Phosphor Evaluation System

1. Introduction:
The Resonance Phosphor Evaluation System uses a computer-controlled tunable vacuum ultraviolet light source with design optimized for test of phosphor emission and luminescence. It delivers vacuum ultraviolet (VUV) excitation wavelengths to samples on a target holder and measures their spectral phosphorescence with a fiber optic coupled CCD spectrometer. This is an excellent system for quality control, fundamental research and development of VUV and UV excited emitting materials.
2. Features:
· 120 to 400 Nm Tunable Light Source with 0.15 to 10 Nm Spectral Bandwidth
· Reference VUV Solar Blind Pmt for Tunable Light Source for Absolute VUV/UV Flux onto Sample and Transmission of Sample
· 10 Sample Positions on Standard Wheel
· Modular Design Allows Change of Emission Sources and Detection Optics
· Compact Footprint less than 1 Sq. Meter
· Turn-Key Oil Free Pumping System
· Emission Spectrometer Uses High Sensitivity UV/Vis/Nir Pmt for a Wide
Spectral Range from 200 To 900 Nm.
· VUV Solar Blind Pmt Transmission Detector
· Order Sorter with 2 Long Pass Filters
· Labview Based Software with USB Interface
3. Specifictions:
Light Source Specifications
|
Specification
|
Minimum
|
Typical
|
Maximum
|
Units
|
Lamp
|
Deuterium
|
Stability: Long Term
|
-
|
1
|
-
|
% per hour
|
Short Term
|
-
|
0.1
|
-
|
% per sec
|
Spectral Output
|
120
|
-
|
400
|
NM
|
VUV Intensity
|
|
2 x 1015
|
5 x 1015
|
Photons/sec/sr
|
UV Intensity
|
|
3 x 1014
|
5 x 1014
|
Photons/sec/sr
|
Certification
|
NIST Traceable Calibration of Irradiance in Vacuum
|
Full Angle Output Cone
|
30
|
45
|
65
|
Degrees
|
Input Power
|
18
|
25
|
30
|
Watts
|
Input Voltage
|
70
|
115
|
260
|
VAC
|
Input Line Frequency
|
50
|
60
|
65
|
Hz
|
Mounting Flange
|
2 3/4 inch Conflat is standard
|
Pulse
|
Modulation to 1 kHz
|
Tunable Monochromator Specifications
|
Specification
|
Minimum
|
Typical
|
Maximum
|
Units
|
Slits Adjustable to
|
10
|
-
|
4000
|
µm
|
Focal Length
|
-
|
200
|
-
|
mm
|
Standard Grating
|
-
|
1200
|
-
|
g/mm
|
Grating Size
|
-
|
42 × 42
|
-
|
mm
|
Resolution
|
-
|
0.1
|
-
|
nm
|
Dispersion
|
-
|
3.45
|
-
|
nm/mm
|
Efficiency (@ 150nm)
|
-
|
> 50
|
-
|
%
|
Wavelength Repeatability
|
-
|
0.2
|
-
|
nm
|
Stray light
|
|
< 2 × 10-4
|
|
|
Vacuum
|
|
< 1 × 10-6
|
|
torr
|
Sample Analysis System Specifications
|
Specification
|
Minimum
|
Typical
|
Maximum
|
Units
|
No. of Samples
|
-
|
10
|
20
|
|
Visible Spectrometer Resolution
|
-
|
1
|
-
|
nm
|
Range
|
200
|
-
|
900
|
nm
|
F No.
|
-
|
3
|
-
|
|
Vacuum
|
-
|
< 1 × 10-6
|
-
|
torr
|
Pumping Speed (to 10-5 torr)
|
|
< 15
|
|
mins
|
|